Kalika Electricals & Electronics

53C

Agronic-53c is compact, light weight, 31/2 digit resistance meter suitable for measurements of resistances of cables, wires, transformers, motor windings, coils and contact resistance of switches, connectors, relays and track resistance of PCB”s etc.

The value of resistance is displayed on 7 segment led display. Four terminal measurement system provides measurements free from errors due to contact resistance and test lead resistances.

Range 1 micro ohms to 19. 99 k ohm in 8 full scale ranges of 1999 micro ohms,

19. 99 milli ohms, 199. 9 milli ohms, 1999 milli ohms, 19. 99 ohms, 199. 9

ohms, 1999 ohms and 19. 99 k ohms.

Resolution 1 micro ohm in 1999 micro ohms range.

0. 01 milli ohm in 19. 99 milli ohms range.

0. 1 milli ohm in 199. 9 milli ohms range.

1 milli ohm in 1999 milli ohms range.

0. 01 ohm in 19. 99 ohms range.

0. 1 ohm in 199. 9 ohms range.

1 ohm in 1999 ohms range

0. 01 k ohms in 19. 99 k ohms ranee.

Accuracy 1 0. 2% + 2 least counts in all ranges except in 200 milli ohms, 20 milli

ohms and 2000 micro ohms ranges.

+- 0. 5% +- 3 least counts in 200 milli ohms, 20 milli ohms and 2000 micro

ohms ranges.

Test current . 0. 1

1 amp dc in 1999 micro ohms, 19. 99 milli ohms, 199. 9 milli ohms and

1999 milli-ohms ranges,

amp dc in 19. 99 ohms range.

1_0 milli amp dc in 199. 9 ohm range.

;1 milli amp dc in 1999 ohms range.

0. 1 milli amp dc in 19. 99 k ohms range.

Display 31/2 digit 7 segment led display with over range indication.
Over-range indication in over-range condition, 1 appears in the left display and the remaining 3

displays are turned off.

Power requirements

Cabinet size

230 volts ac + 10%. 50hzt 10 va.
75 mm h x 185 mm w x 220 mm .
Weight 4 kgs.
Standard Accessories 4 banana to crocodile test leads and instruction manual.
Measurement Range

Brand

Accuracy

Dimension

Weight

Agronic-53c is compact, light weight, 31/2 digit resistance meter suitable for measurements of resistances of cables, wires, transformers, motor windings, coils and contact resistance of switches, connectors, relays and track resistance of PCB”s etc.

The value of resistance is displayed on 7 segment led display. Four terminal measurement system provides measurements free from errors due to contact resistance and test lead resistances.

Range 1 micro ohms to 19. 99 k ohm in 8 full scale ranges of 1999 micro ohms,

19. 99 milli ohms, 199. 9 milli ohms, 1999 milli ohms, 19. 99 ohms, 199. 9

ohms, 1999 ohms and 19. 99 k ohms.

Resolution 1 micro ohm in 1999 micro ohms range.

0. 01 milli ohm in 19. 99 milli ohms range.

0. 1 milli ohm in 199. 9 milli ohms range.

1 milli ohm in 1999 milli ohms range.

0. 01 ohm in 19. 99 ohms range.

0. 1 ohm in 199. 9 ohms range.

1 ohm in 1999 ohms range

0. 01 k ohms in 19. 99 k ohms ranee.

Accuracy 1 0. 2% + 2 least counts in all ranges except in 200 milli ohms, 20 milli

ohms and 2000 micro ohms ranges.

+- 0. 5% +- 3 least counts in 200 milli ohms, 20 milli ohms and 2000 micro

ohms ranges.

Test current . 0. 1

1 amp dc in 1999 micro ohms, 19. 99 milli ohms, 199. 9 milli ohms and

1999 milli-ohms ranges,

amp dc in 19. 99 ohms range.

1_0 milli amp dc in 199. 9 ohm range.

;1 milli amp dc in 1999 ohms range.

0. 1 milli amp dc in 19. 99 k ohms range.

Display 31/2 digit 7 segment led display with over range indication.
Over-range indication in over-range condition, 1 appears in the left display and the remaining 3

displays are turned off.

Power requirements

Cabinet size

230 volts ac + 10%. 50hzt 10 va.
75 mm h x 185 mm w x 220 mm .
Weight 4 kgs.
Standard Accessories 4 banana to crocodile test leads and instruction manual.